Basic Technical Specifications
Measurement range, nT |
> ± 50000 |
Maximal sample rate, Hz |
12 |
Resolution, bits |
16 |
Axial alignment precision (calibrated) |
± 5 min of arc |
Maximal offset, nT |
50 |
Gain error, % of reading |
< 0.1 |
Zero drift over temperature, nT/° C |
< 0.2 |
Gain drift over temperature, % of reading /° C |
< 0.005 |
Operation temperature range, ° C |
-40...+80 |
Power consumption (galvanically insulated power
converter utilizes onboard power supply voltage 24..34V), W |
< 0.4 |
Overall dimensions and weight |
130x80x80 mm, 1 kg |
Shockproof, till, g |
300 |
This instrument is designed for the
highest reliability and uses a structure with redundant equipment in cold reserve mode. It
includes two independent magnetometers with corresponding sensors assembled to form a
small robust monoblock. Special sensor design maintains axial alignment after extreme
shock loads.
A user serial interface transmits output digital signals
into the microsatellite control system. The interface is implemented as
inverse current loop for minimal power consumption and maintains UART
protocol. Both main and redundant channels of communication are galvanically insulated
from other satellite system circuits. |
|