VECTOR MAGNETOMETER
CALIBRATION SYSTEM
LEMI-901
Vector magnetometer Calibration System is intended for the
calibration of different types of vector magnetometers positioned inside the
three-components coil system. Calibration System supports high accuracy regulation and
measurement of currents in the coils and high precision magnetic fields in the magnetic
sensor volume correspondingly, measurement and recording of the corresponding magnetometer
readings for each component both for analog and digital output according to the input data
file, in manual or in automatic mode. The system performs the correction of natural
magnetic field variations as well as temperature correction of own parameters. As a result
of calibration procedure the following parameters are calculated with special software and
printed to the Test Certificate:
| transformation coefficients; |
| angles between the magnetometer components axes; |
| angles between the axes of magnetometer and axes of the
coils; |
| angle and transformation factor correction matrix for
tested magnetometer; |
| zero offsets; |
| linearity errors. |
The special thermal tests subsystem is available intended for the
automatic determination of thermal dependency factors of both electronic and
sensor of the tested magnetometer.
The main parameters
1. Coil system parameters
1.1 |
Coil constants:
X-direction nT/mA 84.8;
Y-direction nT/mA 96.5;
Z-direction nT/mA 73,9.
Accurate values measured after installation. |
1.2 |
Coil orthogonality.
Angles between the magnetic directions measured after installation.
Design accuracy better than ±1 minute of arc in a sphere of 30 cm diameter. |
1.3 |
Coil homogeneity.
Design accuracy is better than 10-5 in a sphere of 30 cm diameter. |
2. Parameters of electronic units
2.1 |
Ranges of currents measurement and control, by each
component |
±2 |
2.2 |
Relative error of currents measurement and control, % |
<0.01 |
2.3 |
Temperature error of reference resistors with temperature
correction in the range (15 - 30)°C, %/°C |
<±0.0005 |
2.4 |
Regulation time for the current in one coil component,
sec |
<2 |
2.5 |
Regulation range of thermal test subsystem, °C
*Note! Negative temperature value depends on
cooling device, not included in the delivery set |
-10... +50 |
2.6 |
Discrepancy between given and real temperature in the
whole range, °C |
±0.5 |
2.7 |
Maximal speed of heating, °C/hour |
10 |
2.8 |
Weight, kg:
control electronics (scaling amplifiers and preamplifiers)
reference resistors 3.5
thermal subsystem 4.0
heater unit 9.5 |
3
3.5
4.0
9.5 |
2.9 |
Power supply |
220V AC, 50Hz |
2.10 |
Power consumption, W |
350 |
|
Publications:
1 |
Vector DC Magnetometers Calibration Error |
IMEKO 2000 XVI WORLD CONGRESS, September 25
28, 2000, Hofburg Austria Vienna |
V.Korepanov
F.Dudkin
K.Pajunpaa |
|
2 |
High-class system for magnetometers
calibration and EMC study test results and development trends |
Proceedings of Int. conf.on metro-logy
Trends and applications in Calibration and testing laboratories (May 16-18,
2000), Jerusalem, Israel, pp. 280-285 |
V.Korepanov
R. Berkman
K.Pajunpaa |
|
3 |
High-class calibration system for DC
magnetometers certification |
Measurement for a sustainable future
(Proceedings of 3rd Biennial Conference MSA99, 22-24 September, 1999, Sydney,
Australia), pp. 59-62, 1999 |
V.Korepanov
K.Pajunpaa
F.Dudkin
Y.Klymovych |
|
4 |
Calibration of ocean bottom magnetometer
|
IUGG-99 Birmingham abstracts (July 19-30,
1999), p. B91
|
V.Korepanov
F.Dudkin |
|
5 |
Calibration system for vector DC
magnetometers |
XIV IMECO World Cong-ress. New
measurements-challenges and visions.
(1-6 June 1997, Tampere, Finland) Vol.IVA, Topic 4, pp.97-102 |
V.Korepanov
K.Pajunpaa
E.Klimovich |
|
|