VECTOR MAGNETOMETER 
CALIBRATION SYSTEM
LEMI-901

VECTOR MAGNETOMETER CALIBRATION SYSTEM

Vector magnetometer Calibration System is intended for the calibration of different types of vector magnetometers positioned inside the three-components coil system. Calibration System supports high accuracy regulation and measurement of currents in the coils and high precision magnetic fields in the magnetic sensor volume correspondingly, measurement and recording of the corresponding magnetometer readings for each component both for analog and digital output according to the input data file, in manual or in automatic mode. The system performs the correction of natural magnetic field variations as well as temperature correction of own parameters. As a result of calibration procedure the following parameters are calculated with special software and printed to the Test Certificate:

bullettransformation coefficients;
bulletangles between the magnetometer components axes;
bulletangles between the axes of magnetometer and axes of the coils;
bulletangle and transformation factor correction matrix for tested magnetometer;
bulletzero offsets;
bulletlinearity errors.


The special thermal tests subsystem is available intended  for the
automatic determination of thermal dependency factors of both electronic and sensor of the tested magnetometer.

The main parameters


1.  Coil system parameters

1.1 Coil constants:
X-direction nT/mA 84.8;
Y-direction nT/mA 96.5;
Z-direction nT/mA 73,9.
Accurate values measured after installation.
1.2 Coil orthogonality.
Angles between the magnetic directions measured after installation.
Design accuracy better than ±1 minute of arc in a sphere of 30 cm diameter.
1.3 Coil homogeneity.
Design accuracy is better than 10-5 in a sphere of 30 cm diameter.

2. Parameters of electronic units

2.1 Ranges of currents measurement and control, by each component ±2
2.2 Relative error of currents measurement and control, % <0.01
2.3 Temperature error of reference resistors with temperature
correction in the range (15 - 30)°C, %/°C
<±0.0005
2.4 Regulation time for the current in one coil component, sec <2
2.5 Regulation range of thermal test subsystem, °C
*Note! Negative temperature value depends on cooling device, not included in the delivery set
-10... +50
2.6 Discrepancy between given and real temperature in the whole range, °C ±0.5
2.7 Maximal speed of heating, °C/hour 10
2.8 Weight, kg:
control electronics (scaling amplifiers and preamplifiers)
reference resistors 3.5
thermal subsystem 4.0
heater unit 9.5

3
3.5
4.0
9.5
2.9 Power supply 220V AC, 50Hz
2.10 Power consumption, W 350

Publications:

1 Vector DC Magnetometers Calibration Error IMEKO 2000 XVI WORLD CONGRESS, September 25 – 28, 2000, Hofburg – Austria – Vienna V.Korepanov
F.Dudkin
K.Pajunpaa

View PDF

2 High-class system for magnetometers calibration and EMC study – test results and development trends Proceedings of Int. conf.on metro-logy “Trends and applications in Calibration and testing laboratories” (May 16-18, 2000), Jerusalem, Israel, pp. 280-285 V.Korepanov
R. Berkman
K.Pajunpaa
View PDF
3 High-class calibration system for DC magnetometers certification Measurement for a sustainable future
(Proceedings of 3rd Biennial Conference MSA99, 22-24 September, 1999, Sydney, Australia), pp. 59-62, 1999
V.Korepanov
K.Pajunpaa
F.Dudkin
Y.Klymovych
View PDF
4 Calibration of ocean bottom magnetometer
IUGG-99 Birmingham abstracts (July 19-30, 1999), p. B91
V.Korepanov
F.Dudkin
 
5 Calibration system for vector DC magnetometers XIV IMECO World Cong-ress. New measurements-challenges and visions.
(1-6 June 1997, Tampere, Finland) Vol.IVA, Topic 4, pp.97-102
V.Korepanov
K.Pajunpaa
E.Klimovich
 